Compressed Air Breathing Apparatus
What is a Breathing Apparatus and where and
why do we use it on board?
A breathing apparatus normally known as the BA set or the
SCBA (Self contained Breathing apparatus) or CABA (Compressed air breathing
apparatus) is an equipment used to supply fresh air for human breathing when a
person is entering any space where the atmosphere is a suspect and may not
support human life.
These include:
·
When
entering into an enclosed space
·
When
entering into a space to fight fire
The Self
Contained Breathing Apparatus (SCBA) or Compressed Air Breathing Apparatus
(CABA):
The equipment is designed on an open-circuit
system. The exhaled air is discharged to
the atmosphere and the wearer of this apparatus gets his fresh air from the air
cylinder attached to the equipment.
The equipment consists of:
·
One or two cylinders containing air
under pressure
·
A harness to mount the cylinders on
the back of the person.
·
A respiratory system which
incorporates a means of reducing the pressure of the air from the cylinder and
of supplying the wearer on demand with air according to his requirements, and
·
A facemask attached to a demand valve,
which maintains a positive pressure inside the mask at all times.
The arrangement of a typical apparatus is as
follows:
·
The cylinders contain approximately
1240 litres of free air compressed to give approximately 31 minutes of full air
usage time and 21 minutes of working duration under normal circumstances. Please note that the usage time greatly
varies from person to person and the workload the person is performing.
·
A reducing valve is connected after
the cylinders to reduce the pressure to approximately about 4 bars. This pressure is further reduced by the
demand valve, which is attached to the mask of the wearer. The demand valve supplies air to the wearer
when he inhales and closes when he exhales.
·
An Exhalation valve releases the
exhaled air from the facemask.
·
When approximately 10 minutes of air
is left (air cylinder pressure about 50 bar), a warning whistle will sound
continuously – warning the user that his air supply would soon be over and he
has to move out. This warning signal
will sound continuously till the air in the cylinder is over.
·
The facemask is made of moulded rubber
with a series of adjustable rubber straps to secure it to the head of the
wearer and fitted with quick release arrangements. The user should have a good field of vision
so that the wearer does not need to turn his head constantly.
·
A gauge is provided to indicate the
pressure of the air in the cylinder.
·
Spare cylinders should be provided for
each set of breathing apparatus and a small air compressor to charge these
cylinders is provided.
If the facemask is dislodged whilst the
wearer is in a smoky or toxic atmosphere, it is vital that the wearer should
hold his breath and immediately replace the dislodged mask. If for any reason, there is likelihood of
delay in doing so, the wearer should be evacuated to fresh air at the earliest,
even though he may not feel any ill effects from the few breaths of
contaminated air, which he has been forced to inhale.
An approximate Consumption and duration
table:
Degrees
of work
|
Air
consumption
(Ltrs/minute)
|
Duration
of cylinder having 1200 ltrs capacity (minutes)
|
Duration
of cylinder having 1800 ltrs capacity (minutes)
|
Resting
|
8
– 12
|
150
– 100
|
225
- 150
|
Light Work
|
12
– 20
|
100
– 60
|
150
– 90
|
Moderate Work
|
20
– 40
|
60
– 30
|
90
– 45
|
Heavy Work
|
40
– 60
|
30
– 20
|
45
– 30
|
Please note in the above table the total full duration is
given. Reduce the safety margin from the
above table to get the working duration.
Nominal
Working Duration = (fully charged
cylinder capacity) minus 10 minutes.
(40)
Here we have taken an average consumption of
40 litres of air per minute being required by human being when doing heavy
work.
Hence if we have a cylinder of 1200 litres
capacity fully charged then
Nominal working duration = 1200/40 = 30
minutes – 10 minutes (period for which the alarm whistle will blow and time
required for the wearer to get back to safe atmosphere) = 20 minutes.
Pre-Operational Checks:
High Pressure Leak Test:
1. On the lung demand valves, press reset lever to switch
‘Off’ positive pressure mechanism.
DO NOT press center of rubber cover while pressing reset
lever, or force lever against stop. See Figure below.
|

|
2. Open cylinder valve slowly, but fully, to
pressurize system.
3. Close cylinder valve and observe pressure
gauge.
4. Gauge reading shall not decrease more than 10 bar in 1
minute.
Note: If a leak is indicated from the demand valve
proceed as follows:
Press center of rubber cover to release
positive pressure mechanism - press reset lever to lift positive pressure
mechanism. Repeat action two or three times to eliminate leak.
If leak not eliminated, do not attempt
repairs on board. Send for service ashore.
Whistle Warning Unit Test:
1. Cover demand valve outlet with ball of
hand, press center of rubber cover (as shown in figure below)

2. Turn ‘On’ positive pressure switch.
3. Slowly vent system by carefully lifting
ball of hand maintaining a slow decrease in pressure.
4. Observe gauge. Whistle should sound at preset
pressure of 55 bar +/- 5bar.
5. Following satisfactory whistle test, press
reset lever to switch ‘Off’ positive pressure.
Should whistle not sound at required pressure, whistle can
be reset using manufacturers instructions AND with Master or Chief Officer
help.
DO NOT ATTEMPT TO ADJUST WHISTLE ON YOUR OWN.
Putting on equipment:
1. Extend shoulder straps and waist belt. Put
on equipment.
2. Fasten buckle (See figure below). Pull ends of waist
belt away from buckle until secure and comfortable on hip. Tuck ends in belt
loops.

3. Pull down shoulder straps (See figure
below) until equipment is secure and comfortable. Tuck strap ends under waist
belt.

Putting on Face piece:
Preparing the mask:
1. Open the straps as far as they will go.

2. Place the carrier strap around your neck.
Fitting the mask:
Beards and sideburns in the sealing area of the facemask
will cause leaks. Therefore the facemask should not be used by anyone with
facial hair of this sort. Spectacle frames in the sealing area also cause
leaks.
Insert chin in chin-piece.
Tighten both lower (1), then upper straps (2)
evenly towards back of head. (See figure below). Only tighten center strap (3)
if necessary.

Connecting Demand Valve to Face piece:
1. Check face piece port and demand valve
O-ring, are clean and undamaged.
2. Push demand valve into face piece port
until it clicks into position.
3. Check attachment is secure. Pull demand valve
away from face piece - there should be no axial movement.
4. On achieving face
seal positive pressure demand valves will automatically activate on first
inhalation.
5. Breathe normally.
6. Before use perform function checks.
Function Checks:
1. Close cylinder valve.
2. Breathe normally to vent system.
3. During venting observe gauge - whistle
alarm should sound at preset pressure of 55bar +/-5bar.
4. When gauge indicates zero, hold breath.
Face piece should hold onto face indicating positive seal.
5. Open cylinder valve slowly, but fully to
pressurize system.
6. Inhale and hold breath. Unit must balance,
i.e. no audible leak.
7. Continue breathing. Expired air should easily
flow out of exhalation valve.
8. Press center of rubber cover on demand valve
checking supplementary supply.
Note: If leak detected, open cylinder valve,
readjust head harness and retest.
During Use:
1. Cylinder pressure must not be less than
80% full.
2. Check gauge reading regularly. Whistle
will sound at 55 bar +/- 5 bar.
3. When whistle sounds, exit and go to safe
area, by shortest and safest route.
4. Do Not remove equipment until in safe
area, clear of hazard.
After Use:
1. Press reset lever to switch ‘Off’ positive
pressure.
2. Press and hold down the button and remove the demand
valve.
3. Removing mask: Insert index finger behind the tabs on
the neck straps and press the buckles forwards with thumb, pulling the mask
forward off your face and then up and back over your head.

4. Close cylinder valve and remove face piece.
5. Unbuckle waist belt, lift shoulder strap
buckles to loosen, remove equipment.
Maintenance on facemask:
Cleaning: The mask must be cleaned after every use. Do not
use any organic solvents such as acetone, alcohol etc.
Clean the mask with a cloth and lukewarm water mixed with
all-purpose detergent. Rinse thoroughly under running water.
Disinfecting: Disinfect mask after every use. Place mask
in disinfectant bath. Only use approved disinfectants - e.g. DAlSYdes. (R54
125)2). Excessive disinfectant concentrations and over-long disinfecting times
can damage the mask. Rinse thoroughly under running water.
Drying: Leave to dry in air. Maximum temperature 60°C.
Keep out of direct sunlight.
Testing: The mask must be tested every time it has been
used, repaired or serviced as per manufacturers instructions.
Visual inspection of the speech diaphragm: The speech
diaphragm and O-ring must be clean and undamaged. If not, clean or replace.


Visual inspection of inhalation valve: The inhalation
valve is visible in the mask connector. Check by blowing into it that both
wings of the valve disc move freely. Replace if seized.

Visual inspection of exhalation valve: Remove the
protective cap from the connector. Press one arm of the spring clamp inwards and
remove the clamp. Hold valve disc by the nipple and lift out of guide. The
valve disc and seat must be clean and undamaged. Clean or replace if necessary.
Moisten valve disc with water and refit. When the seat is
horizontal, the valve disc must slide into the guide under
its own weight. Fit spring clamp and spring so that both clamp arms engage at
the sides. The clamp is marked “L”= left and “R” right and fits at an angle.





Additionally, Maersk requires following checks to be
carried out:
1. Every month the air bottles, bottle valves, reduction
valve, intermediate hose, manometer, carrying back plate, the lung demand
regulator, the warning alarm for withdrawal, and the breathing mask shall be
examined and checked in accordance with the manufacturer's manual.
2. Every year all valves, sealing and speech
membranes shall be checked.
3. Every second year the exhalation disc must
be replaced. This should be done even if the mask has not been used.
4. Every five years the complete breathing
apparatus shall be submitted to a major service check, which should include a
major overhaul of the reduction valves in an authorized workshop and a pressure
test of the air bottles.
5. Pressure test of the air bottles shall be
carried out by an approved service company.
6. If an air bottle shows any sign of
corrosion or if it has been exposed to damage or overheating it shall be
replaced instantly.

Charging Cylinders:
Charge cylinder to pressure stamped on neck,
or shoulder of cylinder.
Refer notes on Breathing Air Compressor.
This exam is good for all the aspiring firemen who want to hone their knowledge when it comes to scba equipment and other stuff.
ReplyDeleteEd of http://www.smokespec.com/